DEL, DUP, and LOH/ROH quality metrics

The following metrics are calculated for CNV (DEL and DUP) and LOH/ROH variants. Available metrics depend on the caller and assay type.

CNVs called by NGS pipeline

VCF Filter

  • The FILTER field value in the VCF, indicating whether a variant passed the caller's quality thresholds or failed specific checks.

  • Requires DRAGEN 4.3+.

Note: For CNVs, the overall quality classification is not determined solely by the VCF FILTER value.

CNVs can be assigned HIGH quality based on CNV-specific criteria (such as SVCLAIM=DJ), even if they do not have a PASS filter in the VCF.

Copy Number

  • Estimated number of copies in the region according to the VCF FORMAT CN field.

Allele-Specific Copy Numbers

  • In addition to the total copy number, based on read depth, the platform can display minor and major allele copy numbers identified by the DRAGEN CNV ASCN module in the format [minor/major].

  • Requires DRAGEN 4.4+.

  • Available for WGS cases from v100.39.

CNV Quality Score

  • Phred-scaled confidence score produced by the CNV caller.

  • Higher values indicate greater confidence.

  • Requires DRAGEN 4.3+.

Size

  • CNV length in base pairs.

  • Size value according to the SVLEN INFO field of the CNV variant.

  • Requires DRAGEN 4.3+.

Bin Count

  • Number of read-depth bins spanning the CNV.

  • Bin count value according to the BC INFO field of the CNV variant.

  • Requires DRAGEN 4.3+.

Mean LRR (Log R Ratio)

  • Shows signal intensity balance across the CNV region.

  • Calculated as log₂(NR).

  • Mean LRR value according to the NR FORMAT field of the CNV/LOH variant.

  • Values close to 0 suggest normal regions; large positive or negative values suggest gains or losses.

  • Requires v100.39 or later.

Mosaic Fraction

  • Provides a mosaic fraction estimate for mosaic events.

  • Mosaic fraction value according to the MF FORMAT field of the CNV/LOH variant.

  • Requires v100.39 or later.

Norm. depth of cov.

  • Normalized depth of coverage according to the NDC FORMAT field, calculated by the DRAGEN Ploidy caller.

  • Requires DRAGEN 4.2+ and v100.39 or later.

Likelihood Ratio

  • Measures confidence in the CNV call. A log₁₀ ratio of ALT versus REF probability.

  • Likelihood ratio value according to the LR INFO field of the CNV variant.

  • Higher values indicate stronger evidence.

  • Available for exome and panel cases.

  • Requires DRAGEN 4.3+.

CNVs and LOH/ROH called by array pipeline

All metrics for CNVs and LOH/ROH called by array pipeline require DRAGEN Array 1.3+.

VCF Filter

  • Indicates whether the variant passed DRAGEN quality checks.

Likelihood Ratio

  • Measures confidence in the CNV call.

  • Higher values indicate stronger evidence.

Mean LRR (Log R Ratio)

  • Calculated as log₂(NR).

  • Shows signal intensity balance across the CNV region.

  • Values close to 0 suggest normal regions.

  • Large positive or negative values suggest gains or losses.

Copy Number

  • Estimated number of copies in the region according to the VCF FORMAT CN field.

Mosaic Fraction

  • Provides a mosaic fraction estimate for mosaic events.

  • Mosaic fraction value according to the MF FORMAT field of the CNV/LOH variant.

CNV Quality

  • Shows the caller’s overall confidence in the variant.

Size

  • CNV length in base pairs.

Bin Count

  • Shows how many data bins (probe regions) support the CNV.

  • A higher bin count means stronger support for the CNV.

LRD

  • Deviation from expected log ratios.

  • Essential for identifying LOH/ROH and assessing call quality.