NGS sample quality

The sample quality indicator rates the sequencing reliability of each sample as High, Moderate, or Low. Sample quality is evaluated using different metrics, depending on the sample file type.

FASTQ sample quality

FASTQ sample quality reflects the confidence level that the sequencing data can support accurate variant calling.

For samples processed from FASTQ files, quality is determined based on the lowest result among these metrics:

FASTQ sample quality evaluation

1

Evaluate sex validation

The system reviews the sex validation result for the sample:

  • If sex validation failed, the system evaluates the FASTQ sample quality as low, and no further metrics are considered.

  • If sex validation passed or is unavailable, the system proceeds to evaluate coverage metrics.

2

Evaluate coverage metrics

The system reviews the results for average coverage and the percentage of target bases with coverage ≥20× and evaluates them as low, moderate, or high using the thresholds in Table 1.

3

Assign quality

The system uses the lower result to assign the FASTQ sample quality.

Table 1. FASTQ sample quality thresholds

Quality
Sex validation
Average coverage
% Target bases with coverage ≥20×

Low

Failed

≤30×

<70%

Moderate

Passed or N/A

30−45×

70−80%

High

Passed or N/A

>45×

>80%

VCF sample quality

The VCF sample quality reflects the confidence level that the sequencing data can support accurate variant interpretation.

For samples processed from VCF files, quality is determined based on:

VCF sample quality evaluation

1

Evaluate sex validation

The system reviews the sex validation result for the sample:

  • If sex validation failed, the system evaluates the VCF sample quality as low, and no further metrics are considered.

  • If sex validation passed or is unavailable, the system proceeds to evaluate error rate.

2

Evaluate error rate

The system reviews the results for error rate and evaluates them as low, moderate, or high using the thresholds in Table 2.

3

Assign quality

The system uses the error rate result to assign the VCF sample quality.

Table 2. VCF sample quality thresholds

Quality
Sex validation
Error rate

Low

Failed

>45%

Moderate

Passed or N/A

25−45%

High

Passed or N/A

<25%