Skip to content
For the complete documentation index, see llms.txt. This page is also available as Markdown.

Lane Splitting and Multi-Analysis by Project

Lane Splitting

The purpose of lane splitting is to enable the reuse of sample indexes (barcodes) on the same flow cell. To accomplish this, the samples indexed with the same barcodes must be physically separated by placing them on different lanes of the flow cell.

Currently, lane splitting functionality with SomaSeq Discovery is only supported on the NovaSeqX platform. To use lane splitting, use the sample sheet to indicate which lane(s) each sample is found in. Recommendation: Edit the "Lanes" column of the SomaSeq Automation System output file(s):

  1. Find the Lanes column in the SomaSeq Automation System output file.

  2. Add lane numbers in which the sample is located as a comma-separated value. Example: '1,2,3,4' for a sample located in lanes 1, 2, 3, and 4.

  3. Upload the modified SomaSeq Automation System output file to the BSSH Run Planner.

Multi-Analysis by Project

The goal of multi-analysis by project is to improve flexibility by enabling multiple analysis outputs from a single sequencing run without needing to requeue the sample sheet. A single project must include at least one sample and can span multiple plates used in a sequencing run. Each project produces one set of output files (including normalized ADATs and a DRAGEN Report).

To use multi-analysis by project, include the project name in the Project column of the SomaSeq Automation System output files. If no project information is provided, DRAGEN Protein Quantification will assume all plates belong to the same project and will assign the project name based on the user provided 'run name'. The multi-analysis by project feature is available for both the NovaSeq 6000 and NovaSeq X platforms.

  1. Find the Project column in the SomaSeq Automation System output file.

  2. Edit Project column in an SomaSeq Automation System output file by assigning a project name to each sample. Project name string should contain only alphanumeric characters and underscores. Not all samples on a plate need to belong to the same project—different project values are permitted across samples within a single plate.

  3. If multiple projects are used on a single plate, leave the Project field blank for control samples (Blanks, Calibrators, QCs). Controls with a blank project value will be shared across all projects on that plate for that KitType.

  4. Upload modified SomaSeq Automation System output file to the Run Planner.

There must be a minimum of 1 sample per project.

Multiple Kit Types on a Single Flowcell

A single flowcell (sequencing run) can contain plates of different kit types (e.g., one Plasma plate and one MultiMatrix plate). Each plate must contain a single kit type with its own controls. During autolaunch, the software performs kit-specific normalization on each plate independently.

If a project includes plates with different kit types or if there are no projects defined and there are multiple kit types on the same flowcell, the software will automatically separate them into separate projects so that each project contains plates of the same kit type. A dataset with two plates of different kit types will produce the same results whether splitting by project or not, as long as each plate is its own kit type.

NOTE: If you want all samples from a sequencing run to be analyzed together, DO NOT include any values in the "Project" column of the sample sheet.

Lane Splitting and Multi-Analysis Examples

Here is an example of what the first lines of two individual SomaSeq Automation System output files may look like after project and lanes values were added. The plate with Barcode "A" adds lanes "1,2,3,4" and project value "ProjectA". The plate with barcode "B" adds lanes "5,6,7,8" and project value "ProjectB".

Last updated

Was this helpful?