Skip to content
For the complete documentation index, see llms.txt. This page is also available as Markdown.

ADAT Content - 9.5k

This page describes the output from the 9.5k SomaSeq Discovery assay, which differs slightly from the 6k assay. Use this for ADATs made on DRAGEN Protein Quantification v2.0 and higher.

The primary output file of DRAGEN Protein Quantification is the ADAT. One ADAT is produced for each normalization step performed. The final ADAT produced (<output_file_prefix>_Step3_SampleNorm.adat) has the normalized counts at the end of the full normalization process.

Supported sample types: Plasma, Serum, CSF, and MultiMatrix (with user-defined tissue types).

There are four key components to the ADAT: the header, SOMAmer metadata, sample metadata, and counts.

ADAT Layout

ADAT Header

Metric
Description

Version

Version of the software used during analysis

Title

User reference for the analysis

SOMAmerReferenceSource

SOMAmer metadata version used during analysis

AssayType

Assay types used

AssayVersion

Version of the SomaSeq Discovery assay (e.g., SomaSeq Discovery v2)

AssayRobot

Liquid handling robot

RunId

Unique identifier for the run (created by sequencing instrument)

Instrument Type

Instrument used for sequencing

Flowcell

Flowcell used for sequencing

YieldDemux

Total number of reads that are demultiplexed

YieldQ30Demux

Total number of reads that are demultiplexed with a passing QC score

Q30WeightedMean

Q30 primary sequencing metric, computed as the Q30 weighted mean across lanes

CreatedDate

Date the run was performed

StudyOrganism

Sample organism (human)

StudyMatrix

Matrix used in study (Serum, Plasma, CSF, or MultiMatrix)

CalibratorId

Lot number(s) of the calibrator samples per plate (JSON format, e.g., {"PlateId":"240320B"})

ProcessSteps

The process steps that occurred to produced the counts in the current ADAT

HybQC_Ref

Reference value for Hyb QC calculation

HybQC_Delta

Threshold delta for HybQC pass/flag determination. If a sample's HybQC value is less than or equal to this threshold, the sample receives PASS.

PlateSOMAmerNormReads_PassFlag

PASS/FLAG for each plate. If greater than 70% of Blank Samples have SOMAmerNormReads_PassFlag = FLAG, PlateSOMAmerNormReads_PassFlag shall be WARNING. Otherwise, it shall have the value "PASS".

PlatformSpecificPlateScale_ScaleFactor

Scale factors for the PlateScale normalization step, when compared to a platform specific reference. The reference is based on data from the sequencing instrument used for the run.

PlatformSpecificCalibrateTailPercent

Percent of Platform Specific Calibration scale factors in tails (outside of the acceptable range of 0.6-1.4) for each plate when compared to a reference. The platform specific reference is based on data from the sequencing instrument used for the run.

PlatformSpecificCalibrateTailPercent_PassFlag

PASS/WARNING for each plate. If PlatformSpecificCalibrateTailPercent for a plate is greater than .15, this value shall be "WARNING". If it is less than or equal to .15, it shall be "PASS".

PlatformSpecificQCCheckTailPercent

Percent of QC scale factors in tails for each plate from comparing Serum QC samples on MultiMatrix plates to their corresponding calibration samples after plate scaling. This is only reported for MultiMatrix.

PlatformSpecificQCCheckTailPercent_PassFlag

PASS/WARNING for each plate. If PlatformSpecificQCCheckTailPercent for a plate is greater than .15, this value shall be "WARNING". If it is less than or equal to .15, it shall be "PASS". This is only reported for MultiMatrix.

CrossPlatformPlateScale_ScaleFactor

Scale factors for the PlateScale normalization step, when compared to a universal reference made with NovaSeq X data.

CrossPlatformCalibrateTailPercent

Percent of Cross Platform Calibration scale factors in tails (outside of the acceptable range of 0.6-1.4) for each plate when compared to a reference. The cross platform reference is based on data from NovaSeq X runs.

QCCheckTailPercent

Percent of QC scale factors in tails for each plate.

QCCheckTailPercent_PassFlag

PASS/FAIL for each plate. If QCCheckTailPercent for a plate is greater than .15, this value shall be "FAIL". If it is less than or equal to .15, it shall be "PASS".

IntraPlateMedianCV

Median Percent CV metric for each plate, computed after Sample Normalization, providing:

  • (sample Standard Deviation of calibrator samples/Mean of calibrator samples)*100

  • (sample Standard Deviation of QC samples/Mean of QC samples)*100

IntraPlateQ75CV

Same as IntraPlateMedianCV (above) but showing 75th percentile of Percent CV instead of median

IntraPlateQ90CV

Same as IntraPlateMedianCV (above) but showing 90th percentile of Percent CV instead of median

InterPlateMedianCV

Median Percent CV metric across plates, providing:

  • (sample Standard Deviation of plate mean count in calibrator samples/Mean of plate mean count in calibrator samples)*100

  • (sample Standard Deviation of plate mean count in QC samples/Mean of plate mean count QC samples)*100

Provided only if there are 4 or more plates in the analysis

InterPlateQ75CV

Same as InterPlateMedianCV (above) but showing 75th percentile of Percent CV instead of median

InterPlateQ90CV

Same as InterPlateMedianCV (above) but showing 90th percentile of Percent CV instead of median

MedianS2B

Median Signal to Background metric for each plate based on human somamers after Plate Normalization, providing:

  • Median Calibrator/Median Blank

  • Median QC/Median Blank

GeneratedBy

Version of SomaData parser used to write ADAT (ex. SomaData_1.0.0)

Sample Metadata

Metric
Description

SampleID

Sample identifier

PlateId

Unique plate identifier

MatrixTubeBarcode

Matrix tube barcode scanned during library prep

BatchID

User-provided batch identifier

InputType

Sample type specified in manifest file.

Examples: Plasma, Serum, CSF, MultiMatrix-<tissuetype>, Blank

MatrixType

Matrix used.

Examples: Plasma, Serum, CSF

SampleType

Type of sample processed.

Examples: Plasma, Serum, CSF, Blank, QC, Calibrator

Control

The control type for the sample. Values: Calibrator, QC, Blank, or empty (for non-control samples).

ControlID

ID of the calibrator, QC, or blank lot (applied to controls only).

KitType

Kit type associated with the sample plate. Values: Plasma, Serum, CSF, or MultiMatrix.

ScriptsVersion

Version of ILASS automation scripts used during library prep (e.g., 3.2.7).

ProbePlate

Probe plate lot number, from library prep

SOMAmerBeadPlate

SOMAmer bead plate lot number. For Plasma/Serum/CSF: SBP1 barcode. For MultiMatrix: MBP barcode.

WellPosition

Location of the sample on the 96 well plate (A1-H12)

Project

(Optional) User-provided project identifier. Multiple projects per plate are supported.

SOMAmerReads

Number of raw counts human SOMAmer reads (excluding control reads)

SOMAmerReads_PassFlag

This flag indicates whether a non-blank sample meets specifications for number of SOMAmer reads. For non-blank samples:

  • SOMAmerReads ≥ 10 million = PASS

  • SOMAmerReads < 10 million = FLAG

Not applied to blank samples.

SOMAmerNormReads

Number of normalized counts human SOMAmer reads at plate scale normalization step (excluding control reads)

SOMAmerNormReads_PassFlag

This flag indicates whether a blank sample meets the specifications for number of normalized SOMAmer reads. For blank samples:

  • SOMAmerNormReads ≤ 20 million = PASS

  • SOMAmerNormReads > 20 million = FLAG

Not applied to non-blank samples.

RefCorr

Spearman correlation value for comparing samples to the plasma, serum, or CSF reference.

HybQC

Hyb QC metric value for the sample, representing the deviation from expected Tm control distribution. Replaces EmpiricalHybTemp from earlier versions.

HybQC_PassFlag

This flag indicates if a sample meets the specification for Hyb QC.

  • HybQC ≤ HybQC_Delta (from ADAT header) = PASS

  • HybQC > HybQC_Delta or missing = FLAG

HybNorm_1_ScaleFactor

The hybridization control scale factor

HybNorm_PassFlag

This flag indicates whether the HybNorm_1_ScaleFactor is within the specified acceptance criteria range of 0.4–2.5. CSF samples (excluding controls) have a different acceptance criteria range of 0.2-5.0.

MedNormInt_<dilution>_ScaleFactor

The MedNormInt scale factor for the dilution group

MedNormInt_PassFlag

This flag indicates whether all dilution group scale factors are within the specified acceptance criteria range of 0.4–2.5 for the MedNormInt step. Applies to blank and calibrator samples.

MedNormExt_<dilution>_ScaleFactor

The MedNormExt scale factor for the dilution group (Plasma/Serum/CSF only; not applicable to MultiMatrix)

MedNormExt_PassFlag

This flag indicates whether all dilution group scale factors are within the specified acceptance criteria range of 0.4–2.5 for the MedNormExt step. CSF samples (excluding controls) have a different acceptance criteria range of 0.125-8.0. Applied to QC and serum/plasma/CSF samples. Not applicable to MultiMatrix.

CrossPlateMedNormInt_all_ScaleFactor

The CrossPlateMedNormInt scale factor specific to MultiMatrix samples

CrossPlateMedNormInt_PassFlag

This flag indicates whether the CrossPlateMedNormInt scale factor is within the specified acceptance criteria range of 0.4–2.5 for MultiMatrix samples

RowCheck_PassFlag

This flag indicates whether the sample's plate failed and if not whether all row scale factors are within the specified acceptance criteria range

SOMAmer Metadata

Metric
Description

SeqId

Unique sequence identifier for the SOMAmer

SeqIdVersion

SOMAmer sequence version

SomaId

SomaLogic-provided identifier

Target

Protein target identifier

Target Full Name

Protein target full name

Type

Target type (protein)

UniProt ID

UniProt identifier

Entrez Gene ID

Entrez Gene identifier

Entrez Gene Symbol

Entrez Gene symbol

HybControl

True if the SOMAmer is a hyb control, else False

Organism

Organism (e.g. Human, Mouse) of the SOMAmer

PlatformSpecificCalibrate_<PlateId>_ScaleFactor

Scale Factor for Platform Specific Calibration

CrossPlatformCalibrate_<PlateId>_ScaleFactor

Scale Factor for Cross Platform Calibration

QCCheck_<PlateId>_ScaleFactor

Scale Factor for QC Check

Dilution

Dilution group classification for the SOMAmer

DRC_Level.<KitType>

Compression of SOMAmer range that occurred during the assay.

References

References used in the analysis. For Plasma/Serum: includes Ref.MedNormExt.{type} and Ref.QCCheck.{type}. Bridging references follow the format Ref.Bridging.<CalibratorId>.<Instrument>.<Flowcell>.v2.

Units

Units of matrix content

Last updated

Was this helpful?